1. Features.
Significant reduction in nitrogen gas use
This model features a high air tight design of the chamber and nitrogen gas flow rate switching mechanism to significantly reduce the amount of nitrogen gas used.
Reduced temperature pull-down time for reduced processing time
This model uses an innovative chamber cooling mechanism to reduce the temperature pull-down time by one-third when compared to conventional models. (Reference value: lowered from 240 minutes to 80 minutes) As a result, specimens can be more quickly removed, and processing time can be reduced.
Profile (example)
2. Specifications.
Model |
IPHH-202L |
Power supply |
200V AC φW 50/60Hz
*Supporting any power supply voltage as an option |
Maximum voltage |
3.8VA |
Method |
Forced hot air circulation |
Performance |
Temperature range |
Ambient temperature +50 to +350°C |
Temperature fluctuation width |
±0.1°C at 100°C/±0.2°C at 200°C
±0.2°C at 350°C |
Temperature uniformity |
±0.5°C at 100°C/±1.5°C at 200°C
±3.0°C at 350°C |
Temperature heat-up time |
Within 80 minutes from ambient temperature to +350°C |
N2 gas |
Gas |
N2 gas (ambient temp. dry gas) |
Min. oxygen concentration |
100 ppm or less (at 60 to 100 NL/min) |
Gas introduction |
Mouthpiece ring joint 3/8 type pipe fitting |
Inside capacity |
216 L |
Dimensions |
Inside dimensions |
W600×H600×D600mm (excluding protrusions) |
Outside dimensions |
W1190×H1700×D825mm (excluding protrusions) |
Weight |
250 kg |
Instrumentation |
Operating mode |
Constant operation, program operation |
Setting and indication ranges |
Temperature setting: 0 to +360°C
Time: 0 to 99 hours and 59 minutes, 100 to 999 hours
(time set in hour increments over 100 hours) |
Setting and indication resolution |
Temperature: 1°C
Time: 1 minute |
Indication accuracy |
Temperature: within ± (2°C + 1 digit)
Time: ±300 ppm of read value |
* Performance is without a specimen at an ambient temperature of +23°C.
* The model is for operational purposes and may be changed after order. Thank you for your understanding.