Surface Resistivity
|
103~1013(Ω/sq. or Ω)
|
Mode
|
① Standard Mode(Ohm per square, Surface Resistivity)
② Mat surface - Mat resistance measurement Mode (Ohms, Resistivity to Ground)(w/ cable)
③ IEC Standard measurement Mode follow the IEC standard probe (surface + weight)
|
LCD display
|
Exponent display (one digit beyond the decimal point)
Remaining measurement time by bar-graph indicator
Remaining battery value
|
LED display
|
① TEST
Green LED indicator: Standard Mode (Ohm per square, Surface Resistivity)
Red LED indicator: Mat-Ground resistance measurement Mode (Ohms, Resistivity to Ground) and IEC standard measurement Mode
② Measurement voltage (10V,100V)
③ CHARGE
Red LED indicator: in process of charging
|
Measurement time
|
10sec, 30sec, 60sec
|
Range
|
Can be set up the max and min of measurement value (by 0.5 exponent)
|
Accuracy
|
±0.5 decade (ex.108.5±0.5)
|
Measurement volatage
|
Automatically adjust/chamge
① Cube of 3 to 6 : 10V, 10mA max
② Cube of 6 to 13 : 100V, 1mA max
|
Source
|
Inside battery
|
External outlet
|
① External probe connector
② AC adaptor connector (battery charging USB mini type)
|
Measurement are
|
50mm (probe length)×50mm (probe distance) (in measurement)
|
Probe material
|
Special conductive rubber (volume resistivity 0.02Ω・㎝ max.)
|
Environment
|
15 ℃~ 25 ℃, under 60%R.H.
|
Dimension
|
114(W)×78(D)×69(H) max. (mm)
|
Case
|
Stainless steel
|
Weight
|
body 700g
|