Differential Pressure - Piezoresistive
|
Measuring range
|
0 to +300 hPa
|
Accuracy
|
±0.5 hPa (0.0 to 50.0 hPa)
±1 % of mv (50.1 to 100.0 hPa)
±1.5 % of mv (Remaining Range)
|
Resolution
|
0.1 hPa
|
Flue gas O₂
|
Measuring range
|
0 to 21 Vol.%
|
Accuracy
|
±0.2 Vol.%
|
Resolution
|
0.1 Vol.%
|
Reaction time t₉₀
|
< 20 s
|
Flue gas CO (with H₂-compensation)
|
Measuring range
|
0 to 8000 ppm
|
Accuracy
|
±10 ppm or ±10 % of mv (0 to 200 ppm)
±20 ppm or ±5 % of mv (201 to 2000 ppm)
±10 % of mv (2001 to 8000 ppm)
|
Resolution
|
1 ppm
|
Reaction time t₉₀
|
< 60 s
|
Flue gas COlow
|
Measuring range
|
0 to 500 ppm
|
Accuracy
|
±2 ppm (0 to 39.9 ppm)
±5 % of mv (40 to 500 ppm)
|
Resolution
|
0.1 ppm
|
Reaction time t₉₀
|
< 40 s
|
Flue gas NO
|
Measuring range
|
0 to 3000 ppm
|
Accuracy
|
±5 ppm (0 to 100 ppm)
±5 % of mv (101 to 2000 ppm)
±10 % of mv (2001 to 3000 ppm)
|
Resolution
|
1 ppm
|
Reaction time t₉₀
|
< 30 s
|
Flue gas Nolow
|
Measuring range
|
0 to 300 ppm
|
Accuracy
|
±2 ppm (0 to 39.9 ppm)
±5 % of mv (40 to 300 ppm)
|
Resolution
|
0.1 ppm
|
Reaction time t₉₀
|
< 30 s
|
Flue gas Draught
|
Measuring range
|
-9.99 to +40 hPa
|
Accuracy
|
±0.02 hPa or ±5 % of mv (-0.50 to +0.60 hPa)
±0.03 hPa (+0.61 to +3.00 hPa)
±1.5 % of mv (+3.01 to +40.00 hPa)
|
Resolution
|
0.01 hPa
|
Temperature
|
Measuring range
|
-40 to +1200 °C *
|
Accuracy
|
±0.5 °C (0 to +100.0 °C)
±0.5 % of mv (Remaining Range)
|
Resolution
|
0.1 °C (-40 to +999.9 °C)
1 °C (> +1000 °C)
|
* Measuring range depends on the connected flue gas probe
Flue gas degree of effectivity, Eta (calculated)
|
Measuring range
|
0 to 120 %
|
Resolution
|
0.1 %
|
Flue gas loss (calculated)
|
Measuring range
|
0 to 99.9 %
|
Resolution
|
0.1 %
|
Flue gas CO₂ calculation (calculated from O₂)
|
Measuring range
|
0 to CO₂ max (Display range)
|
Accuracy
|
±0.2 Vol.%
|
Resolution
|
0.1 Vol.%
|
Reaction time t₉₀
|
< 40 s
|
Pressure measurement
|
Measuring range
|
0 to +300 hPa
|
Accuracy
|
±0.5 hPa (0.0 to 50.0 hPa)
±1 % of mv (50.1 to 100.0 hPa)
±1.5 % of mv (Remaining Range)
|
Resolution
|
0.1 hPa
|
Flue gas CO (without H₂-compensation)
|
Measuring range
|
0 to 4000 ppm
|
Accuracy
|
±20 ppm (0 to 400 ppm)
±5 % of mv (401 to 2000 ppm)
±10 % of mv (2001 to 4000 ppm)
|
Resolution
|
1 ppm
|
Reaction time t₉₀
|
< 60 s
|
Ambient CO
|
Measuring range
|
0 to 500 ppm
|
Accuracy
|
±5 ppm (0 to 100 ppm)
±5 % of mv (> 100 ppm)
|
Resolution
|
1 ppm
|
Reaction time
|
Approx. 35 s
|
with CO probe 0632 3331
Ambient CO₂
|
Measuring range
|
0 to 1 Vol.%
0 to 10000 ppm
|
Accuracy
|
±75 ppm or ±3 % of mv (0 to 5000 ppm)
±150 ppm or ±5 % of mv (5001 to 10000 ppm)
|
Reaction time
|
Approx. 35 s
|
with ambient CO₂ probe 0632 1240
Gas leak measurement for combustible gases (via gas leak
detection probe)
|
Measuring range
|
0 to 10000 ppm CH₄ / C₃H₈; Display range
|
Accuracy
|
Signal optical display (LED) audible signal via buzzer
|
Reaction time t₉₀
|
< 2 s
|
with gas leak detection probe 0632 3330
General technical data
|
Dimensions
|
270 x 90 x 65 mm
|
Operating temperature
|
-5 to +45 °C
|
Protection class
|
IP40
|
Warranty
|
Probe: 48 months; Gas sensors (O₂, CO) 48
months; NO sensor 24 months; COlow sensor 24 months; Thermocouple 12 months;
rechargeable battery: 12 months
|
Display size
|
240 x 320 pixels
|
Display function
|
Colour graphic display
|
Power supply
|
Rechargeable battery pack 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2
A
|
Maximum
|
500,000 readings
|
Storage temperature
|
-20 to +50 °C
|
Weight
|
600 g (without rechargeable battery)
|