Temperature - NTC
|
Measuring range
|
-40 to +150 °C
|
Accuracy
|
±0.2 °C (-10 to +50 °C)*
0.4 °C (Remaining Range)*
|
Resolution
|
0.1 °C
|
*Accuracy information applies only to instrument without probes
connected
Temperature - TC Type K (NiCr-Ni)
|
Measuring range
|
-200 to +1370 °C
|
Accuracy
|
±0.4 °C (-100 to +200 °C)*
1 °C Remaining Range*
|
Resolution
|
0.1 °C
|
*Accuracy information applies only to instrument without probes
connected
Differential pressure (internal sensor) - Piezoresistive
|
Measuring range
|
0 to 100 hPa
|
Accuracy
|
±0.2 % of fsv
|
Resolution
|
0.01 hPa
|
Static pressure
|
1000 hPa (abs)
|
Overload
|
300 hPa
|
Zeroing
|
to 2.5 hPa
|
Differential pressure (external probe) - Piezoresistive
|
Measuring range
|
0 to 2000 hPa
|
Accuracy
|
±0.1 % of mv*
|
Resolution
|
0.1 Pa (0638 1347)
0.01 hPa (0638 1547)
0.001 hPa (0638 1447)
0.1 hPa (0638 1847 / 0638 1647)
|
*Accuracy information applies only to instrument without probes
connected
Differential pressure (external probe) - Ceramic
|
Measuring range
|
-1 to 400 bar
|
Accuracy
|
±0.2 % of fsv*
|
Resolution
|
0.01 bar
|
*Accuracy information applies only to instrument without probes
connected
General technical data
|
Dimensions
|
219 x 68 x 50 mm
|
Operating temperature
|
0 to +50 °C
|
Housing
|
ABS
|
Channels
|
2 external; 1 internal
|
Warranty
|
2 years
|
Measuring rate
|
0.04 s
|
Connection
|
Hose: inner Ø 4 mm outer Ø 6 mm
|
Battery type
|
9 V (6LR61)
|
Battery life
|
Continuous operation w/ internal pressure sensor: 30 h; With
rech. battery: 10 h; With carbon battery: 18 h
|
Display type
|
LCD
|
Display function
|
7 segment display and point matrix
|
Display update
|
2x per second, in fast measurement 4x per second
|
Power supply
|
Battery/Rechargeable battery,Mains unit 12 V
|
Interface
|
RS232
|
Memory
|
100 kB; 25,000 measuring values
|
Storage temperature
|
-20 to +70 °C
|
Weight
|
300 g
|