Model |
SK-L200TII(temperature measurement type)
|
Display range |
-45 to 205.0°C
(depending on a probe connected) |
Resolution |
0.1°C at -9.9 to 205.0°C
1°C at -10°C or less |
Accuracy |
±(0.1°C + 1 digit) |
Probes connected |
SK-LTII series probes |
Sampling (display) |
About 1 sec. |
Display functions |
TIME (display time), INT (log interval) REC (during logging), RES (during preset standby), OUT (during real-time transmission), LCD bars (used memory)
CALL (call for the current measurement data during logging)
Current temperature, Current humidity, Current date and time, low battery symbol, Temp. upper limit (Hi), Temp. lower limit (Lo), Temp. alarm ON (Alarm), logging option, setting value of upper and lower limits, error message, ID number |
Memory capacity |
8100 lines of data, but depending on the logging option
(four logging potions: Once, Repeatability, Ends at, By page) |
Log interval |
1sec., 2sec., 5sec., 10sec., 15sec., 30sec.
1min., 2min., 5min., 10min., 15min., 30min., 60min., 90min. |
Communication |
USB |
Operation ambient |
-10 to 60°C, no condensing |
Storage ambient |
0 to 50°C |
Power requirement |
"AAA" size alkaline battery x 2 |
Battery life |
Under continuous measurement mode : more than 500 hours
Under logging mode: possible to log 8100 data with every log interval
(ex.: about 506 days with log interval of 90 minutes.) |
Dimensions |
(W) 100 x (H) 73 x (D) 22 mm |
Weight |
About 123 g (exclusive of batteries) |
Materials |
ABS resin |
Accessories |
CD-Rom (Data analysis and real-time software), USB cable (1.5m) , USB port cap (factory set), hand strap, "AAA" size battery x 2 |