Specifications of HD-1000 series
Model |
HD-1100
HD-1150 |
HD-1400
HD-1450 |
HD-1500
HD-1550 |
Waterproof specifications
(Equivalent to IPX5) |
◯ |
◯ |
— |
Display |
LCD |
LED |
LCD with Backlight |
Input connector |
ASP model: HD-1*0* / ANP model: HD-1*5* |
Channel |
1ch |
Input type |
Thermocouple Input: Type E, K (HD-1*0*) Not switchable
Thermocouple Input: Type E, K, J, T or R (HD-1*5*) Not switchable |
Signal source resistance |
MAX 500Ω |
Measurement Range |
Table 1 |
Measurement Accuracy |
Table 2 |
Reference junction
compensation accuracy |
±0.2°C at 25°C±10°C [±0.4°F at 77°F±18°F] |
Temperature Coefficient |
±0.01% of F/S per °C [±0.01% of F/S per °F] (F/S : full scale) |
Environment |
Operation |
0 to 40°C, 0 to 80%RH (Non condensing)
32 to 104°F, 0 to 80%RH (Non condensing) |
Storage |
-20 to 50°C, 0 to 85%RH (Non condensing)
-4 to 122°F, 0 to 85%RH (Non condensing) |
Battery life |
300 h |
150 h |
150 h |
Power supply |
Dry battery |
4 AA Alkali Batteries |
AC power |
— |
— |
AC-Adaptor (Option) |
Sampling rate |
approx.300ms |
approx.300ms*1 |
Plinearlizer |
Digital Linearizer method (Compliant with IEC 60584-1 (1995)) |
Dimensions |
approx. 76(w)×167(H)×36(D) mm [Protruded parts excluded] |
Weight |
approx. 350g [Include Batteries] |
Supplied accessories |
— |
— |
Communication Cable
Software (AMS-100)
(Table 4) |
Analog Output |
— |
— |
— |
Alarm Output |
— |
— |
— |
Memory function |
— |
— |
◯ (Table 3) |
°C / °F Change |
— |
— |
— |
Compatible standards |
CE Marking, RoHS |
*1 Sampling period will be a user-selected time interval when time interval is specified at 10 seconds or more.
Table 1 : Measurement Range
Thermocouple Type |
°C |
1°C Resolution |
0.1°C Resolution |
E |
-200 to 800°C |
-100.0 to 500.0°C |
K |
-200 to 1370°C |
J |
-200 to 1100°C |
T |
-200 to 400°C |
-100.0 to 400.0°C |
R |
0 to 1760°C |
— |
Thermocouple Type |
°F |
1°F Resolution |
0.1°F Resolution |
E |
-328 to 1472°F |
-148.0 to 932.0°F |
K |
-328 to 2498°F |
J |
-328 to 2012°F |
T |
-328 to 752°F |
-148.0 to 752.0°F |
R |
32 to 3200°F |
— |
* Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.
Table 2 : Measurement Accuracy
Type |
°C |
E, K, J, T |
0.1°C Resolution |
0°C to |
± (0.05% of reading + 0.2°C) |
to 0°C |
± 0.5°C |
1°C Resolution |
0°C to |
± (0.1% of reading +1°C) |
to 0°C |
± (0.5% of reading +1°C) |
R |
1°C Resolution |
± (0.1% of reading +3°C) |
Type |
°F |
E, K, J, T |
0.1°F Resolution |
32°F to |
± (0.05% of reading + 0.4°F) |
to 32°F |
± 0.5°F |
1°F Resolution |
32°F to |
± (0.1% of reading + 2°F) |
to 32°F |
± (0.5% of reading +2°F) |
R |
1°F Resolution |
± (0.1% of reading +6°F) |
Table 3 : Memory function
Time interval |
second:1, 5, 10, 30 minutes: 1, 5, 10, 30, 60 and manual memory |
Memory capacity |
9999 data |
Table 4 : Operational environment OS software
OS |
Microsoft® Windows® 7, Windows® 8.1, Windows® 10 |
Memory · Hard-disk space |
Normal operating environment of OS |
* All PCs under the recommended environment are not guaranteed to work successfully with AMS-100.
* AMS-100 can be used only by the user that has a system administrative right (Adoministrator)
* AMS-100 does not support Mac OS.
* Microsoft® Windows® 7, Windows® 8.1, Windows® 10, and Excel® are registered trademarks of Microsoft Corporation in the United States and other countries.